辐射力天平Radiation Force Balances
CW标准声源Check Sources
超声功率计Ultrasound Power Meter
高精度测量超声功率值,广泛实用于医学超声的诊断和治疗。
满足IEC超声功率测量标准。
辐射力天平RFB CW标准声源 超声功率计
辐射力天平 用于医学超声诊断和治疗的高精度功率测量。为了满足美国FDA 510(K)、IEC61157和IEC61689的要求,必须对超声功率源的功率值进行精确测量。辐射力天平RFB可以对1MHz-10MHz,10mW-100W的功率进行精准测量。两种靶可选,可以单独测量也可以与计算机配套使用。
CW标准声源 是为了定期对辐射力天平RBF和超声功率计进行规范校准,以确保功率测量仪器自身的准确性而满足国家测量标准。
射力天平RFB
Precision Acoustics RFB’s is for a highly specified model suitable for the majority of medical diagnostic and physiotherapy ultrasound equipment: Using the suspended target the power generated by HIFU transducers can be measured
◇Calibrated frequency range:1MHz to 10MHz*.
◇Power range 20mW to 100W+ but the calibration range is 20mW to 10W*.
◇Overall measurement uncertainty* – for 50mW to 5W, less than +/-7% at 95% confidence.
◇For power levels greater than 100mW the typical type A (random) measurement uncertainties is typically
only 1%.
◇Maximum transducer diameter for absorbing well target 5cm* but the suspended target will accommodate large transducers -contact us with your specific requirements.
*The RFB may be used outside of this range with increased uncertainty levels
Listed are the options available for NPL calibrations which are traceable to the UK’s Primary Standard.
◇Calibration at 1 and 3MHz, Powers: 0.1W, 0.5W, 1W, 5W and 10W.
◇Calibration at single frequency at Powers to be agreed.
◇Calibration at 1 and 3MHz, Powers: 0.1W, 0.5W, 1W, 5W and 10W AND 3 Outputs in the range 10mW to
1W at 0.928MHz and 3.151MHz AND 2 Outputs in the range 10mW to 200mW at 5.287MHz and
7.415MHz and 9.541MHz.
CW标准声源
单探头校正标准声源NPLCS_SINGLE 双探头校正标准声源NPLCS_DUAL
激励Excitation | 在的频率可以两个连续波输出 |
探头Transducer | 一或两个探头1,2,3.5,5,7,10MHz可选 直径均为19mm,除了1MHz的为23mm |
标称功率 Nominal Power Output | 20mW、100mW或500mW 前面板可选 |
稳定性Stability | 使用1000小时内高于+/- 2.5% |
校正精度Calibration accuracy | 80-250V,50-60Hz |
输入Electrical input | +/- 7% 作为标准 |
超声功率计UPM DT-1000PA
功率范围Power Range | 0-30W |
分辨率Resolution/Uncertainly | 1mW(0.1mg in grams mode) |
显示灵敏度Display Sensitivity | 0.001W |
精确度Accuracy | +/- 3%@10W,25℃ |
归零模式Zeroing Method | 自动 |
响应时间Response Time | <3.0秒 |
探头尺寸Max Transducer Size | 直径4 1/2” |
探头操作频率Transducer Operating Freq. | 500KHz-10MHz |
测试介质Test Media | 除气水Degassed Water |
计算机接口Computer Interface | RS-232,600-19200 Baud |
默认波特率Default Baud Rate | 2400 |
尺寸Size | H11″*L15″W10″ |
重量Weight | 16LBS |